Quick answer
SpO₂ estimates the percentage of hemoglobin oxygen-binding sites that are occupied, using a pulse oximeter. PaO₂ is the partial pressure of dissolved oxygen in arterial blood, measured with an arterial blood-gas sample. They are related, but they are not interchangeable.
What does each number measure?
Pulse oximetry uses light absorption to estimate arterial hemoglobin oxygen saturation noninvasively. It is useful for following trends and detecting many oxygenation problems, but it is not a direct measurement of PaO₂ or tissue oxygen delivery.
PaO₂ describes the partial pressure of oxygen dissolved in arterial plasma. It affects the movement of oxygen into blood and the degree of hemoglobin saturation. The relationship is shaped by the sigmoid oxyhemoglobin dissociation curve.
Why is the relationship nonlinear?
At relatively high PaO₂, hemoglobin is already close to fully saturated. Increasing PaO₂ further may cause little change in SpO₂. At lower PaO₂, the curve is steeper, and a relatively small fall in oxygen tension may be associated with a much larger fall in saturation. Changes in pH, temperature and carbon dioxide can shift the dissociation curve, further affecting the relationship.
This is also why a saturation reading does not tell the whole story about oxygen carried by blood. Most arterial oxygen content is bound to hemoglobin; anemia can substantially reduce oxygen content despite a saturation near normal.
When can pulse oximetry mislead?
Readings may be affected by poor peripheral perfusion, movement, nail products and device performance. The FDA also highlights accuracy differences related to skin pigmentation and other patient or measurement factors. A discordance between the displayed saturation and the clinical picture calls for reassessment rather than blind trust in the number.
A blood gas can provide PaO₂ and other physiological information that a conventional pulse oximeter does not supply, such as PaCO₂ and acid–base measurements.
Clinical interpretation and important limitations
A pulse oximeter estimates the fraction of hemoglobin binding sites occupied by oxygen, while PaO2 measures dissolved oxygen tension. Anemia, dyshemoglobins, low perfusion and other factors may alter what these tests tell us; neither result alone establishes tissue oxygen delivery.
Common questions
Can PaO₂ change while SpO₂ stays nearly the same?
Yes, especially on the relatively flat upper portion of the dissociation curve.
Can SpO₂ be normal in severe anemia?
Yes. Saturation does not measure hemoglobin concentration.
Does a normal SpO₂ exclude a ventilation problem?
No. Pulse oximetry does not directly measure CO₂ elimination.